Chromatographic Symptoms
3
Troubleshooting
25
Larger Peaks Than Expected
•
Check each configured column’s dimensions against the
actual column dimensions. (See “Configurable Items to
Always Keep Current” .)
•
Check the autosampler injection volume.
•
Check the vial caps.
•
Check configured syringe size. Some syringe sizes are
specified at half-capacity. If the maximum syringe volume is
marked at half-height on the barrel, not at the top of the
barrel, enter
twice
the labeled volume when configuring the
syringe size.