Checking for Leaks
7
Troubleshooting
79
To Check for External Leaks
Check for leaks at these connections:
•
Gas supply bulkhead fittings
•
Gas cylinder fitting
•
Regulator fittings
•
Traps
•
Shut-off valves
•
T-fittings
Perform a pressure drop test.
1
Turn off the GC.
2
Set the regulator pressure to 415 kPa (60 psi).