FID Ignitor Does Not Glow During Ignition Sequence
1
Remove the detector top cover.
2
Turn the FID flame
On.
3
Observe the ignitor plug though the FID chimney. The small
hole should glow during ignition sequence.
If the test fails, check for the following possible causes:
•
The ignitor may be bad; replace the ignitor.
•
Detector temperature is set to < 150 °C. Agilent recommends
operating the FID at > 300 °C.
•
The ignitor is not making a good connection to the ground:
•
The ignitor must be tightly screwed into the FID castle
assembly.
•
The three T-20 Torx screws that hold the collector
assembly in place must be tight.
•
The brass knurled nut that holds the FID castle assembly
in place must be tight.
Perform FID maintenance if these parts are corroded or
oxidized.