30
Troubleshooting
3
Chromatographic Symptoms
Peak Boiling Point or Molecular Weight Discrimination Poor
If you have trouble with peak boiling point or molecular weight
discrimination (inlet discrimination), do the following:
•
Check the inlet for contamination. Clean and change the
liner if necessary. Replace all inlet consumable parts. See the
Maintenance manual.
•
Adjust the inlet temperature.
•
Run standards against a known method to determine
expected performance.
For any inlet operating in split mode with any detector
•
Check liner type.
•
Increase the inlet temperature and verify that the insulation
cup is installed and contains insulation.
•
Check column cut and installation into the inlet. See the
topic for the SS, PTV, and VI.
For any inlet operating in splitless mode with any detector
•
Check the inlet for leaks. (See “Checking for Leaks” .)
•
Check liner type.
•
Verify that the oven starting temperature is less than the
solvent boiling point.
•
Check column cut and installation into the inlet See the topic
for the SS, PTV, and VI.
•
Check that the solvent vapor volume does not exceed the
liner capacity.
•
Check for appropriate purge delay time.