GC Not Ready Symptoms
4
Troubleshooting
59
Cannot Set a Flow or Pressure
If you cannot set a flow or pressure using the split/splitless,
PTV, VI, or cool on-column inlets, do the following:
•
Check the column mode.
•
Check that a capillary column is configured to the correct
inlet.
•
Check the configured column dimensions.
•
Check that the flow is turned on.
If you cannot set a flow or pressure using the purged packed
inlet, do the following:
•
Check the column mode. The packed inlet uses
flow
control
for undefined columns.
•
Check that the flow is turned on.