Chromatographic Symptoms
3
Troubleshooting
23
Peak Areas Not Repeatable
•
Check the ALS syringe operation. (See the Troubleshooting
section of the 7683B Automatic Liquid Sampler Installation,
Operation and Maintenance manual.)
•
Replace the syringe.
•
Check for leaks in the inlet, liner (as applicable), and column
connection. (See “Checking for Leaks” .)
•
Check sample level in vials.
•
Run replicates of known standards to verify the problem.
•
Consider if this is this first run. (Has the GC stabilized?)