Oscilloscope Mode - The "Measure" Menu
SCOPIX Oscilloscopes
IV - 81
Oscilloscope Mode (cont'd)
Unattached manual
cursors
Used for linking or not linking the manual measurement cursors (1 and 2)
to the reference trace.
When the "Unattached cursors" menu is selected, cursors 1 and 2 can
be moved freely over the whole screen.
•
The "
"
symbol indicates that the "Unattached cursors" menu is active.
•
To deactivate this menu, deselect it by pointing with the stylus.
In the case of "Automatic measurements" and manual measurement
activation:
If …
then …
… the
manual
cursors and
the
automatic
markers are
displayed together,
… the
automatic
measurements
are performed on the portion of the
trace defined between the
manual
cursors.
… the portion defined
between the
manual
cursors
is too small [in this case, the
fixed markers (+) will not be
displayed],
… the
automatic
measurements
selected are impossible, so "-.--" is
indicated in the measurement
display area.
Deselect the automatic measurements to validate the manual
measurements (dt, dv).
Summary of Contents for OX7102-CII
Page 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Page 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Page 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Page 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Page 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Page 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Page 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Page 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Page 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Page 183: ...Applications SCOPIX Oscilloscopes IX 183...
Page 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...