Recorder Mode - The “Trigger” Menu
Recorder Mode (cont’d)
The "Trig" Menu
Triggering
Selection of trigger type and level on each channel. Triggering takes place if a
condition described by a line of the “Trigger” table is verified.
The trigger level should be defined in the channel measurement dynamic.
Source
Indicates the channel number.
Type
The type of trigger for each channel.
There are several possible types:
¾
None
¾
Lower than
¾
Higher than
¾
Lower/Higher
¾
Exterior
In "Recorder" mode, several conditions are monitored simultaneously.
Level 1
The main trigger threshold level can be set using the stylus.
Level 2
The auxiliary trigger threshold level can be set using the stylus. This tab is
enabled only if external trigger type is selected.
min duration
The fault will be evidenced if the fault condition defined by the type and
thresholds is present for a parameterizable period from 3,2 ms to 670 ks
according to the used time base (from 160 µs to 670 ks if the l’ EXTENDED
ACQUISITION MEMORY option).
*
Trigger levels are entered into the current value display area after modification.
Fine adjustment is possible.
SCOPIX Oscilloscopes
VII - 131
Summary of Contents for OX7102-CII
Page 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Page 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Page 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Page 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Page 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Page 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Page 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Page 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Page 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Page 183: ...Applications SCOPIX Oscilloscopes IX 183...
Page 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...