Oscilloscope Mode - The "Measure" Menu
SCOPIX Oscilloscopes
IV - 79
Oscilloscope Mode (cont'd)
Measurement
conditions
•
The measurements are performed on the displayed part of the trace.
•
Any change to the signal will lead to updating of the measurements.
They are refreshed in step with acquisition.
•
To perform automatic measurements on specific portions of the signal,
frame the required measurement area with the manual cursors so
that the markers show the new location of the automatic measurement.
•
Measurement precision is optimum if two complete signal periods are
displayed.
Automatic
measurement
presentation
T = 1/F
W+
W-
Vmax
Vhigh
Vlow
Vamp Vpp
>5%T
Trise
Tfall
0%
10%
50%
90%
100%
>5%T
Vmin
t0
t1
t2
t3
t4
t5
t6
Vavg
•
Positive overshoot = [100 * (Vmax – Vhigh)] / Vamp
•
Negative overshoot = [100 * (Vmin – Vlow)] / Vamp
•
Vrms
=
[
(y
y
) ]
1
n
i
i 0
i n
GND
2 1/2
=
=
∑
−
•
Vavg
=
1
n
(y
y
)
i
i 0
i n
GND
=
=
∑
−
•
Vsum =
∑
=
=
δ
×
n
i
0
i
)
(y
t
i
Y
GND
= value of the point representing zero volt
low
high
P
avg
W
W
Trise
Tfall
Summary of Contents for OX7102-CII
Page 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Page 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Page 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Page 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Page 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Page 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Page 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Page 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Page 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Page 183: ...Applications SCOPIX Oscilloscopes IX 183...
Page 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...