Recorder Mode - The “Trigger” Menu
Recorder Mode (cont’d.)
Display
(fault capture,
file capture)
Two cases arise:
•
the horizontal zoom is enabled,
•
the horizontal zoom is disabled.
Horizontal zoom
disabled
Modification of display:
•
The cursors are no longer displayed.
•
The measurements displayed do not have the same significance.
•
All the measurements are related to the current recording.
•
“Trigger time” and “Sample value at trigger time” measurements are only
displayed if the trigger has taken place.
•
Maximum and minimum searches cannot be enabled (see §. Display Menu,
p. 137).
•
The automatic measures cannot be enabled (see §. Measurement Menu,
p. 139).
Horizontal zoom
enabled
The display is identical to normal mode.
•
The measurements displayed have the same significance as in normal mode
(see §. Display, p. 137).
•
Maximum and minimum searches are possible (see §. Display Menu, p. 137).
•
Automatic measurements can be enabled (see §. Measurement Menu, p. 139).
Acquisition
time of last
acquired
sample
Trigger time
Acquisition time
of first viewable
sample
Value of
last
acquired
sample
Current
recording
faults
Value of
sample at
time of
trigger
SCOPIX Oscilloscopes
VII - 135
Summary of Contents for OX7102-CII
Page 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Page 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Page 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Page 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Page 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Page 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Page 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Page 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Page 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Page 183: ...Applications SCOPIX Oscilloscopes IX 183...
Page 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...