Oscilloscope Mode - The Keys
IV - 36
SCOPIX Oscilloscopes
Oscilloscope Mode (cont’d)
4 "Trigger" keys
Sets the trigger
level
to the average value of the signal (50%) without
modifying the trigger coupling.
When pressed in combination with a
CHx
key, this activates the same
function, after first selecting the corresponding channel as the trigger
source.
Selects the trigger
gradient
(up
or down
) by successive presses.
The gradient is indicated in the status area.
Successive presses can be used select one of the following
acquisition
modes
:
Single shot
(Mono) = SINGLE
Triggered
Trig
Automatic
(Auto) = REFRESH
•
"
SINGLE" mode
:
A single acquisition is authorized, triggered by pressing the
RUN HOLD
key
.
For any further acquisition, the trigger circuit must be reset by pressing
the RUN HOLD key.
If the time base is less than 100 ms/div
SINGLE mode can be
accessed via the "Mono (< 100 ms/div)" option.
If not, Single mode can be accessed via the “Roll Mono (> 50
ms/div.)” option and ROLL mode is automatically activated.
•
"
TRIGGERED
"
mode:
The screen's content is only refreshed when there is a trigger event
linked to the signals present on the oscilloscope's inputs (CH1, CH2,
CH3, CH4).
If there is no trigger event linked to the signals present on the inputs (or
if there is no signal on the inputs), the trace is not refreshed.
•
"
AUTOMATIC
"
mode
:
The screen's content is refreshed even if the trigger level is not detected
on the signals present on the inputs.
When there is a trigger event, screen refreshing is managed as in the
"Triggered" mode.
•
Allows acquisition to be started and stopped in "TRIGGERED" and
"AUTOMATIC"
modes.
•
Resets the trigger circuit in "SINGLE" mode.
Acquisition is initiated according to the conditions defined by the acquisition
mode (
SGLE REFR
key).
The acquisition status is indicated in the status area:
RUN
= started
STOP
= stopped
PRETRIG
= acquisition
Summary of Contents for OX7102-CII
Page 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Page 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Page 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Page 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Page 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Page 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Page 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Page 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Page 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Page 183: ...Applications SCOPIX Oscilloscopes IX 183...
Page 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...