Applications
Applications
1. Display of the
calibration signal
•
Using the key shown
opposite
, select the "Oscilloscope" mode.
•
Connect the
Probix
adapter of a 1/10-ratio HX0030 probe to input CH1.
A message indicating the characteristics of the probe is displayed briefly,
confirming that it has been detected.
The
Probix
menu (Vert
CH1
Probix
) can be used to configure buttons
A
and
B
of the probe.
•
Use this menu to assign button
A
to increasing the sensitivity and button
B
to reducing it by selecting: Sensi/-.
See the description in the chapter on Probix p. 17.
•
The same menu can be used to modify the colour of CH1 trace
Red
•
Via the probe (with its earth) connect the calibrator output to input CH1
(Probe Adjust:
≈
3 V,
≈
1 kHz) located on the side of the instrument.
Connect the cold point of the 1/10 probe to the cold point of the
calibration output of the probes.
•
Check that the coefficient of the 1/10 probe has been taken into account:
Vert Menu
CH1
Vertical Scale
Coefficient: 10.
The sensitivity and the measurements take the probe's coefficient into
account.
•
Validate the signal:
Vert
Menu Display
Trace 1
or by pressing the CH1 key
or on the display of the CH1 trace parameters.
•
Adjust the CH1 sensitivity:
Vert Menu
CH1
Sensitivity/coupling: 500 mV/div. (1/10 probe)
or by using buttons
A
and
B
of the
HX0030
probe
or the keys
opposite.
•
Adjust the CH1 coupling:
Vert Menu
CH1
Sensitivity/coupling
AC
or by pressing the AC/DC GND key.
•
Adjust the scan speed:
using the scrollbar in the time base window: 500 µs/div
or the keys
opposite.
•
Select the trigger parameters:
Trig. Menu
Parameter
Principal
Source: CH1
Coupling: AC
Front: + (or using the key
opposite).
•
Adjust the trigger mode:
Trig. Menu
Automatic mode
or using the SGLE REFR key.
•
Use the RUN HOLD key to start acquisition ("RUN" mode)
SCOPIX Oscilloscopes
IX - 157
Summary of Contents for OX7102-CII
Page 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Page 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Page 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Page 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Page 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Page 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Page 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Page 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Page 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Page 183: ...Applications SCOPIX Oscilloscopes IX 183...
Page 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...