Technical Specifications
X - 186
SCOPIX Oscilloscopes
Technical Specifications
(
cont'd)
« Oscilloscope » Function
Trigger circuit
Characteristics
OX 7042 - OX 7062
OX 7104 - OX 7102
Trigger
sources
CH1, CH2, CH3, CH4
(
OX 7104
)
CH1, CH4
(
OX 7102
)
Trigger
mode
Automatic
Triggered
Single shot
Auto Level 50%
AC:
BW 10 Hz to 200 MHz
D
C:
BW 0 to 200 MHz
HFreject: BW 0 to 10 kHz
LFreject:
BW 10 kHz to 200 MHz
Trigger
coupling
without bandwidth
limit
With bandwidth limitation activated, the bandwidth is limited to 20 MHz.
Trigger
gradient
Falling or rising
Trigger
sensitivity
0.6 div. at 1 kHz (noise rejection mode inactive)
Noise
rejection
≈
±1.5 div.
Trigger
level
Variation range
±10 div.
on edge
on pulse width < t
≈
t
> t
from 20 ns to 20 s
Trigger after delay of 120 ns to 20 s
qualifier source: CH1 (CH2) (CH3) CH4
trigger source: CH1 (CH2) (CH3) CH4
Trigger after counting 3 to 16,384 events
qualifier source: CH1 (CH2) (CH3) CH4
counting source: CH1 (CH2) (CH3) CH4
trigger source: qualifier or counting source
Trigger
type
TV on CH1 only:
- Selection of line number and polarity, with 525 lines (PAL) and 625 lines
(SECAM), even or odd line field
- TV trigger sensitivity: > 1 div.
HOLDOFF
Adjustable from 160 ns to 30 sec.
Summary of Contents for OX7102-CII
Page 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Page 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Page 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Page 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Page 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Page 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Page 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Page 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Page 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Page 183: ...Applications SCOPIX Oscilloscopes IX 183...
Page 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...