Applications
Applications (cont’d)
If necessary:
•
Modify the trigger level using the stylus:
- by moving the symbol T (Trigger) on the screen. The value of the trigger
level is indicated in the bottom right-hand corner of the screen.
- or through the trigger parameters menu:
Trig. Menu
Parameter
Principal
Level
•
Modify the vertical positioning of the curve:
- by using the stylus to move symbol 1, on the left of the screen.
- or by using the keys
opposite.
The key opposite is used to make these adjustments automatically.
HX0030 probe
compensation
Adjust the low-frequency compensation of the probe so that the signal plateau is
horizontal (see figure below).
Over-compensated probe
Compensated probe
Under-compensated probe
Adjust the screw on the Probix HX0030 probe to adjust the compensation.
IX - 158
SCOPIX Oscilloscopes
Summary of Contents for OX7102-CII
Page 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Page 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Page 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Page 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Page 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Page 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Page 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Page 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Page 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Page 183: ...Applications SCOPIX Oscilloscopes IX 183...
Page 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...