Oscilloscope Mode - The "Trigger" Menu
IV - 66
SCOPIX Oscilloscopes
Oscilloscope Mode (cont'd)
Count
Selects the edge trigger with counting of events.
The count is triggered by the auxiliary source. The main source serves as a
clock for the count.
Effective triggering occurs after the end of the count on the next event from the
main source.
Trigger delay
3
Adjustment with the stylus using the setting scroll bar to choose the
number of events required.
Double-tapping in this field displays a virtual numeric keypad which can be
used to directly input the value.
Holdoff
232µs
Adjustment with the stylus using the setting scroll bar, disabling of
the trigger for a predefined period and, among other things,
stabilization of the trigger on pulse trains.
Double-tapping in this field displays a virtual numeric keypad which can be used
to directly input the value.
Auxiliary source
selects a
channel
as a trigger source.
Coupling
Selection of the filter for the auxiliary trigger source:
AC
AC coupling (10 Hz to 200 MHz):
blocks the DC component of the signal
DC
DC coupling (0 to 200 MHz):
allows the entire signal through
LF Reject
Reject of source signal frequencies < 10 kHz
facilitates observation of signals with a DC component
HF Reject
Rejection of source signal frequencies > 10 kHz:
facilitates observation of signals with high-frequency noise
Edge
Selection of the trigger gradient:
+
trigger on ascending edge
-
trigger on descending edge
The selected trigger edge is indicated the status area.
Level
1.09V Adjustment of the trigger level with the stylus on the scroll bar.
Noise reject
•
No:
hysteresis
≈
0.5 div.
•
Yes:
introduces a hysteresis of
≈
1.5 div.
Summary of Contents for OX7102-CII
Page 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Page 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Page 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Page 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Page 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Page 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Page 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Page 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Page 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Page 183: ...Applications SCOPIX Oscilloscopes IX 183...
Page 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...