Recorder Mode - “Display” Menu
Recorder Mode (cont’d)
Faults
The search for faults enables successive recordings in “fault capture” or “file
capture” (memory) mode to be explored (.REC).
All files with a “.REC” extension are analysed and each fault is displayed.
When one of these faults is selected, it is displayed on the screen. The main
cursor is fixed at the fault’s trigger point and the auxiliary cursor is placed at
the right of the screen: the window is accessible from: Display
Faults.
Various information is provided in this window:
•
Title
:
total number of faults contained in the memory
•
No
:
fault number
•
Acq. date/time
: date (day, month) and time of fault acquisition
•
Source
:
symbol for type of trigger used for this fault and source
•
File
:
name of file in which the fault is recorded
The term “memory” in filenames means that the fault captured is in
auxiliary storage. If a file is called anything other than “memory”, all the
faults in the “memory” file will be lost.
Example
Following a file capture, the “Faults” option will allow the different recordings to
be analyzed:
Full screen
Switches from normal display mode to “full screen” display mode and vice
versa.
The display is organized so as to leave the biggest surface area possible for
curve plotting: only the permanent settings remain, as well as the option to
display the automatic measurement window.
•
The “full screen” function has the same effect as the key shown opposite.
•
The “
”
symbol indicates that the full screen mode is enabled.
This function can also be accessed by double-tapping the trace display area
with the stylus.
VII - 138
SCOPIX Oscilloscopes
Fault n°11 in file
selected:
4214ae6e.REC. Its
triggering source is
outside window.
It was captured on
17 February at
14:47:08.
Display of fault n°11
with the main cursor
on the trigger point
and the auxiliary
cursor on the last
sample acquired.
The recorder er is in
memory display. To
return to normal
display, deselect
Recall ‘.REC’ in the
Memory menu.
The symbol
M
is a
reminder that
the recorder
is displaying
a memory
(.REC).
Summary of Contents for OX7102-CII
Page 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Page 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Page 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Page 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Page 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Page 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Page 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Page 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Page 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Page 183: ...Applications SCOPIX Oscilloscopes IX 183...
Page 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...