Oscilloscope Mode - The "Trigger" Menu
SCOPIX Oscilloscopes
IV - 63
Oscilloscope Mode (cont'd)
Pulse
Selection of pulse-width trigger. In all cases, the effective trigger occurs on the
pulse trailing edge.
<
triggers on a pulse if its duration is less than the value set
=
triggers on a pulse if its duration is equal to the value set
>
triggers on a pulse if its duration is greater than the value set
25.0 µs Adjustment with the stylus using the time setting scroll bar
Double-tapping in this field displays a virtual numeric keypad which can be used
to directly input the value.
The choice of the
or
edge in the "Main" tab defines the limits of the
analysis:
•
edge
defines a pulse between
and
•
edge
defines a pulse between
and
Example
Signal injected on CH1: 1 pulse of 50 µs (repetitive or not)
The trigger parameters in the main menu are active (Source, Level, Edge, etc.).
The oscilloscope is triggered when the signal's pulse width is equal to the
specified pulse width (25.0 µs + tolerance).
The trigger occurs on the ascending edge of the pulse, but is effective on the
trailing edge.
Summary of Contents for OX7102-CII
Page 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Page 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Page 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Page 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Page 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Page 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Page 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Page 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Page 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Page 183: ...Applications SCOPIX Oscilloscopes IX 183...
Page 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...