Applications
Applications (cont’d)
•
Select the trigger mode:
Trig.
Menu Automatic mode
or using the SGLE REFR key.
•
Use the RUN HOLD key to start acquisition (RUN mode)
or use the time base menu.
The acquisition status (Ready, RUN, STOP) is indicated on the right, under
Reminder
er status display zone.
the display of the trace, in the trigg
•
Optimize the time base speed to observe several complete TV lines.
Example of a video signal
Use the manual cursors to check the duration of a line
IX - 162
SCOPIX Oscilloscopes
•
the key
opposite.
•
To move the cursors freely, select:
M
sors.
•
Use the stylus to position cursors 1 and 2 on the start and end of the
duration of a line
Display the manual cursors:
Menu Measure
Manual measurements (dt, dv)
or using
easure
Menu Unattached cur
signal, respectively.
The measurements between the 2 cursors are indicated under the curve
display.
Example: dt = 64.00 µs =
Summary of Contents for OX7102-CII
Page 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Page 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Page 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Page 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Page 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Page 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Page 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Page 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Page 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Page 183: ...Applications SCOPIX Oscilloscopes IX 183...
Page 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...