
Recommended procedures for each of the other SETUP menu items are
as follows:
1. Set Block Bin #
This menu item displays the current block probe bin setting and
allows the setting to be changed. However, the block probe bin
setting should only need to be changed when a new block probe of
a different bin number is installed. In that case, make sure the new
block probe bin number is recorded in the service log at the end of
this user’s
guide and proceed as follows.
NOTE
Changing the bin setting requires recalibration of
the instrument.
At
“Set Block Bin #”
, press the START button. The display will
report the current block bin setting. Compare the current block bin
setting with the new block bin number recorded in the service log.
If the current block bin setting is the same as that recorded in the
service log, simply press the STOP button to return to
“Set Block
Bin #”
. If the current bin number is not the same as the new bin
number, enter the required bin number by means of the numeric
keypad.
Press ENTER to store the new setting (or STOP to restore the orig-
inal setting).
Recalibrate the instrument as instructed in Chapter 4.
21
Installation & Setup
Summary of Contents for 4C3
Page 1: ...The Advanced Cryoscope Model 4C3 User s Guide Part No 4C35 Rev17 042318 ...
Page 6: ...vi The Advanced Model 4C3 User s Guide Notes ...
Page 10: ...x The Advanced Model 4C3 User s Guide Notes ...
Page 14: ...Notes xiv The Advanced Model 4C3 User s Guide ...
Page 68: ...46 The Advanced Model 4C3 User s Guide Notes ...
Page 116: ...94 The Advanced Model 4C3 User s Guide Notes ...
Page 122: ...100 The Advanced Model 4C3 User s Guide Notes ...
Page 126: ...104 The Advanced Model 4C3 User s Guide Notes ...
Page 132: ...110 The Advanced Model 4C3 User s Guide Notes ...
Page 136: ...114 The Advanced Model 4C3 User s Guide Date Problem Symptom Action ...