Installation & Setup
In order to set up your instrument properly, it is important that
you read and follow the steps in this section. Please follow these
steps carefully and be sure to read
Chapter 2 — Instrument
Operation
before attempting to run tests on your instrument.
A. Things to do in advance
1.
Prepare Bench Space.
The area beneath your instru-
ment should be kept clear and adequate space must be
provided around the back and sides to allow un-imped-
ed fan-driven air circulation.
NOTE
Bright light shining directly into the cassette
deck area can interfere with the operation of
the instrument.
2.
Locate and Test Grounded Outlet.
A three-prong
grounded (earthed) 100-250V, 50-60Hz outlet capa-
ble of continuously supplying 2.5 amperes at 250V to
5 amperes at 100V is required within five feet of the
instrument.
WARNING
This instrument must be properly
grounded (earthed). If the instrument is
not grounded properly, its operation will
be impaired and a safety hazard may
exist. It is not enough to simply plug the
instrument into a grounding outlet. Have
the outlet tested. Record the results in the
service log at the end of this guide.
1
Summary of Contents for 4C3
Page 1: ...The Advanced Cryoscope Model 4C3 User s Guide Part No 4C35 Rev17 042318 ...
Page 6: ...vi The Advanced Model 4C3 User s Guide Notes ...
Page 10: ...x The Advanced Model 4C3 User s Guide Notes ...
Page 14: ...Notes xiv The Advanced Model 4C3 User s Guide ...
Page 68: ...46 The Advanced Model 4C3 User s Guide Notes ...
Page 116: ...94 The Advanced Model 4C3 User s Guide Notes ...
Page 122: ...100 The Advanced Model 4C3 User s Guide Notes ...
Page 126: ...104 The Advanced Model 4C3 User s Guide Notes ...
Page 132: ...110 The Advanced Model 4C3 User s Guide Notes ...
Page 136: ...114 The Advanced Model 4C3 User s Guide Date Problem Symptom Action ...