![Advanced Instruments 4C3 User Manual Download Page 104](http://html.mh-extra.com/html/advanced-instruments/4c3/4c3_user-manual_2856032104.webp)
82
The Advanced
®
Model 4C3 User’s Guide
“
Thermoelectrics
Driver Failure”
“Illegal Error
Code”
“Reset Probe
Configuration”
displayed
during the
power-up
sequence,
followed by
“Select Test
Item
b. Press > to advance to
“Set Sample Bin #” in
the SETUP menu and
reset the sample probe
bin # (see Chapter 1,
Section K.2).
c. Reset the stir ampli-
tude, as necessary, to
the number recorded
in the service log in
the back of the user’s
guide.
Switch the instrument
off, then on.
Retry the test.
If the error message
persists, see Chapter 5,
Section C for further
assistance.
“Cooling
System Error”
Switch the instrument
off, then on.
Retry the test.
If the error message
persists, see Chapter 5,
Section C for further
assistance.
Switch the instrument
off, then on.
Retry the test.
If the error message
persists, see Chapter 5,
Section C for further
assistance.
Symptom
Check
Suggested Action
Summary of Contents for 4C3
Page 1: ...The Advanced Cryoscope Model 4C3 User s Guide Part No 4C35 Rev17 042318 ...
Page 6: ...vi The Advanced Model 4C3 User s Guide Notes ...
Page 10: ...x The Advanced Model 4C3 User s Guide Notes ...
Page 14: ...Notes xiv The Advanced Model 4C3 User s Guide ...
Page 68: ...46 The Advanced Model 4C3 User s Guide Notes ...
Page 116: ...94 The Advanced Model 4C3 User s Guide Notes ...
Page 122: ...100 The Advanced Model 4C3 User s Guide Notes ...
Page 126: ...104 The Advanced Model 4C3 User s Guide Notes ...
Page 132: ...110 The Advanced Model 4C3 User s Guide Notes ...
Page 136: ...114 The Advanced Model 4C3 User s Guide Date Problem Symptom Action ...