![Advanced Instruments 4C3 User Manual Download Page 40](http://html.mh-extra.com/html/advanced-instruments/4c3/4c3_user-manual_2856032040.webp)
J. Optional barcode scanner
In conjunction with a commercially-available barcode scanner, the
4C3 can provide automatic identification of each sample on the
printed results as indicated in Chapters 1 and 2. A D-Type, 15-pin
barcode port (see figure 7) is provided in the back of the 4C3 for
connecting and providing power to such a device. For proper oper-
ation, the barcode port requires a 1200 bps, TTL-level signal pro-
viding asynchronous serial data containing 1 start bit, 8 data bits, 1
stop bit and no parity. The barcode port pin out is shown in table 4
on next page.
A suitable barcode scanner is available from Advanced Instruments.
To interface with the 4C3, the barcode scanner must be pro-
grammed as follows, referring to the scanner users guide:
1200 bps
CR suffix
disable beep after good decode
triggerless trigger mode (optional)
18
The Advanced
®
Model 4C3 User’s Guide
Signal Pin Direction
+5V DC
1
to reader
receive data
10
from reader
gnd/earth 9 common
Table 4:
Barcode Port Connections
Summary of Contents for 4C3
Page 1: ...The Advanced Cryoscope Model 4C3 User s Guide Part No 4C35 Rev17 042318 ...
Page 6: ...vi The Advanced Model 4C3 User s Guide Notes ...
Page 10: ...x The Advanced Model 4C3 User s Guide Notes ...
Page 14: ...Notes xiv The Advanced Model 4C3 User s Guide ...
Page 68: ...46 The Advanced Model 4C3 User s Guide Notes ...
Page 116: ...94 The Advanced Model 4C3 User s Guide Notes ...
Page 122: ...100 The Advanced Model 4C3 User s Guide Notes ...
Page 126: ...104 The Advanced Model 4C3 User s Guide Notes ...
Page 132: ...110 The Advanced Model 4C3 User s Guide Notes ...
Page 136: ...114 The Advanced Model 4C3 User s Guide Date Problem Symptom Action ...