Troubleshooting Table
“Result = 0,
Calibration
Okay?”
“Standards
Reversed?”
“Test Time-out
Error”
“Recalibration
Needed”
Check the calibration
(unless running a 0 mOsm
sample).
Recalibrate unless running
a 0 mOsm sample.
Calibrate the instrument.
If the error message repeats
after successful re-calibra-
tion, see Chapter 5, Section
C for further assistance.
Recalibrate with the correct
calibration standards.
Did you introduce the
right calibration standards
in the right order?
1. Was the sample OK?
2. Is the block probe bin
# correct?
3. Check the alignment
by running the Head
Up/Down Test (see
Chapter 5, Section
G.8).
a. Prepare another sample
and repeat the test.
b. Check/correct the block
probe bin #.
c. If the error message
persists, see Chapter 5,
Section C for further
assistance.
“Calibration
Out of Range”
See Symptom item
“Results not repeatable...”.
Recalibrate.
Procedure, technique.
“No Plateau”
a. Test another aliquot of
that sample in a clean
sample tube.
b. If the error message
persists, see Chapter 5,
Section C for further
assistance.
Symptom
Check
Suggested Action
Summary of Contents for 4C3
Page 1: ...The Advanced Cryoscope Model 4C3 User s Guide Part No 4C35 Rev17 042318 ...
Page 6: ...vi The Advanced Model 4C3 User s Guide Notes ...
Page 10: ...x The Advanced Model 4C3 User s Guide Notes ...
Page 14: ...Notes xiv The Advanced Model 4C3 User s Guide ...
Page 68: ...46 The Advanced Model 4C3 User s Guide Notes ...
Page 116: ...94 The Advanced Model 4C3 User s Guide Notes ...
Page 122: ...100 The Advanced Model 4C3 User s Guide Notes ...
Page 126: ...104 The Advanced Model 4C3 User s Guide Notes ...
Page 132: ...110 The Advanced Model 4C3 User s Guide Notes ...
Page 136: ...114 The Advanced Model 4C3 User s Guide Date Problem Symptom Action ...