92
The Advanced
®
Model 4C3 User’s Guide
Symptom
Check
Suggested Action
“Calibrator
Out of Range”
“Insufficient
Data”
“No Sample
Tube”
“WARNING-
Sample Count
Incorrect”
Sample tested during
autocal was outside
expected range.
Repeat or use manual
calibration.
Number of samples
tested/completed, do not
match number of samples
input for batch or barcode
run.
Enter a lower number of
runs.
Verify that the number of
runs requested in recall
do not exceed number of
runs processed.
a. Check printout for test
errors or cancellations
that may have thrown
off count.
b. If completed tests and
errors add up correctly,
disregard.
“Serial Loop
Failed”
Review procedure. A
sample tube is required
for the function
requested.
If the error message
persists, see Chapter 5,
Section C for further
assistance.
Procedure.
1. Is serial loop-back
connector attached?
2. Did you wire loop-
back connector cor-
rectly?
3. Try lower baud rate.
a. Attach connector.
b. See Chapter 5, Section
G.4.
c. Change baud rate per
Chapter 1, Section K.8.
Summary of Contents for 4C3
Page 1: ...The Advanced Cryoscope Model 4C3 User s Guide Part No 4C35 Rev17 042318 ...
Page 6: ...vi The Advanced Model 4C3 User s Guide Notes ...
Page 10: ...x The Advanced Model 4C3 User s Guide Notes ...
Page 14: ...Notes xiv The Advanced Model 4C3 User s Guide ...
Page 68: ...46 The Advanced Model 4C3 User s Guide Notes ...
Page 116: ...94 The Advanced Model 4C3 User s Guide Notes ...
Page 122: ...100 The Advanced Model 4C3 User s Guide Notes ...
Page 126: ...104 The Advanced Model 4C3 User s Guide Notes ...
Page 132: ...110 The Advanced Model 4C3 User s Guide Notes ...
Page 136: ...114 The Advanced Model 4C3 User s Guide Date Problem Symptom Action ...