acumed Ankle Plating System 3 Manual Download Page 26

PKGI-79-F Effective 03-2019

 

 

English – EN / PAGE 26 

 

EN 

STORAGE INSTRUCTIONS: 

Store in a cool dry place 

and keep away from direct sunlight. Prior to use, inspect 
product package for signs of tampering or water 
contamination. Use oldest lots first. 
 

APPLICABILITY: 

These materials contain information 

about products that may or may not be available in any 
particular country or may be available under different 
trademarks in different countries. The products may be 
approved or cleared by governmental regulatory 
organizations for sale or use with different indications or 
restrictions in different countries. Products may not be 
approved for use in all countries. Nothing contained on 
these materials should be construed as a promotion or 
solicitation for any product or for the use of any product in 
a particular way which is not authorized under the laws 
and regulations of the country where the reader is 
located. 
 

FURTHER INFORMATION: 

To request further 

material, please see the contact information listed on this 
document. 

SYMBOL LEGEND 

 

Consult instructions for use 

 

Caution 

 

Sterilized using ethylene oxide 

 

Sterilized using irradiation 

 

Use-by date 

 

Catalogue number 

 

Batch code 

 

Authorized representative in the European 

Community 

 

Manufacturer 

 

Date of manufacture 

 

Do not resterilize 

 

Do not re-use 

 

Upper limit of temperature

 

Caution: For Professional Use Only. 

Summary of Contents for Ankle Plating System 3

Page 1: ...l 11 rue Emile ZOLA BP 2332 38033 GRENOBLE CEDEX 2 FRANCE 33 4 76 86 43 22 Acumed LLC 5885 NE Cornelius Pass Road Hillsboro OR 97124 9432 1 503 627 9957 acumed net 2797 Ankle Plating System 3 PKGI 79...

Page 2: ...NTENTS Click on a language to navigate to the page English US 3 Dansk DA 11 English EN 19 Deutsch DE 27 EL 36 Espa ol ES 46 Fran ais FR 55 Italiano IT 64 Nederlands NL 73 Norsk NO 82 Portugu s PT 90 S...

Page 3: ...plantation Patients who are unwilling or incapable of following postoperative Ankle Plating System 3 care instructions are contraindicated for these devices These devices are not intended for screw at...

Page 4: ...f this implant These cautions include the possibility of the device or treatment failing as a result of loose fixation and or loosening stress excessive activity or weight bearing or load bearing part...

Page 5: ...cal allergic or adverse foreign body reaction resulting from implantation of a foreign material may occur Nerve or soft tissue damage necrosis of bone or bone resorption necrosis of the tissue or inad...

Page 6: ...id scratching the surface Mechanical Processing Equipment Washer disinfector neutral enzymatic cleaner or neutral detergent with a pH 8 5 Instrument Cleaning Acumed Instruments and Accessories must be...

Page 7: ...of twenty 20 minutes Use a nylon soft bristled brush to gently scrub instruments until all visible debris is removed Pay special attention to hard to reach areas Pay special attention to any cannulat...

Page 8: ...3 Remove instruments from enzyme solution and rinse in deionized water for a minimum of one 1 minute 4 Place instruments in a suitable washer disinfector basket and process through a standard washer...

Page 9: ...to steam sterilization and sterility assurance in health care facilities Consult your equipment manufacturer s written instructions for specific sterilizer and load configuration instructions Gravity...

Page 10: ...t be approved for use in all countries Nothing contained on these materials should be construed as a promotion or solicitation for any product or for the use of any product in a particular way which i...

Page 11: ...e Ankelskinnesystem 3 TIL OPERATIONSKIRURGEN OG ASSISTERENDE SUNDHEDSPERSONALE Dansk DA SIDE 11 DA instruktioner efter operation Disse anordninger er ikke beregnet til skruefastg relse eller fiksering...

Page 12: ...e advarsler omfatter risikoen for at anordningen eller behandlingen fejler som et resultat af l s fiksering og eller l sning belastning for stor aktivitet v gtst tte eller trykbelastning is r hvis imp...

Page 13: ...ler p f ring af for stort tryk p implantatet under indf ring Der kan forekomme implantatmigration og eller l sning Der kan forekomme overf lsomhed over for metal histologiske eller allergiske reaktion...

Page 14: ...antatet omhyggeligt med destilleret eller renset vand Anvend destilleret eller renset vand til den sidste skylning 4 T r implantatet med en ren bl d fnugfri klud for at undg at ridse overfladen Mekani...

Page 15: ...pl sninger n r de eksisterende opl sninger bliver markant kontaminerede 2 L g instrumenterne i en enzymatisk opl sning til de er helt d kkede Aktiver alle bev gelige dele for at lade reng ringsmidlet...

Page 16: ...enter Brug af en spr jte eller vandstr le vil forbedre skylning af sv rt tilg ngelige omr der og t tsiddende pasflader 3 Tag instrumenterne op af enzymopl sningen og skyl dem i deioniseret vand i mind...

Page 17: ...10 Comprehensive guide to steam sterilization and sterility assurance in health care facilities Omfattende vejledning i dampsterilisering og sterilitetssikring p sundhedsinstitutioner Se producentens...

Page 18: ...eller restriktioner i forskellige lande Produkterne er muligvis ikke godkendt til brug i alle lande Intet i disse materialer m fortolkes som reklame eller salgsfors g for et produkt eller for brug af...

Page 19: ...plantation Patients who are unwilling or incapable of following postoperative Ankle Plating System 3 FOR THE PERSONAL ATTENTION OF THE OPERATING SURGEON AND SUPPORTING HEALTHCARE PROFESSIONALS English...

Page 20: ...of this implant These cautions include the possibility of the device or treatment failing as a result of loose fixation and or loosening stress excessive activity or weight bearing or load bearing pa...

Page 21: ...antation of a foreign material may occur Nerve or soft tissue damage necrosis of bone or bone resorption necrosis of the tissue or inadequate healing may result from the presence of an implant or due...

Page 22: ...id scratching the surface Mechanical Processing Equipment Washer disinfector neutral enzymatic cleaner or neutral detergent with a pH 8 5 Instrument Cleaning Acumed Instruments and Accessories must be...

Page 23: ...of twenty 20 minutes Use a nylon soft bristled brush to gently scrub instruments until all visible debris is removed Pay special attention to hard to reach areas Pay special attention to any cannulat...

Page 24: ...e 3 Remove instruments from enzyme solution and rinse in deionized water for a minimum of one 1 minute 4 Place instruments in a suitable washer disinfector basket and process through a standard washer...

Page 25: ...T79 2010 Comprehensive guide to steam sterilization and sterility assurance in health care facilities Consult your equipment manufacturer s written instructions for specific sterilizer and load config...

Page 26: ...s in different countries Products may not be approved for use in all countries Nothing contained on these materials should be construed as a promotion or solicitation for any product or for the use of...

Page 27: ...er Materialempfindlichkeit Wenn eine Materialempfindlichkeit vermutet wird sollten vor der Implantation entsprechende Tests durchgef hrt werden Patienten die zur Befolgung der postoperativen Pflegema...

Page 28: ...n vertraut zu machen Au erdem liegt es in der Verantwortung des Chirurgen sich mit den relevanten Ver ffentlichungen vertraut zu machen und das Verfahren vor der Anwendung mit erfahrenen Kollegen zu b...

Page 29: ...gepr ft werden Die Implantate sind vor Kratzern und Kerben zu sch tzen Derartige Spannungskonzentrationen k nnen zum Versagen f hren Ein mehrmaliges Biegen der Platten kann das Produkt schw chen und...

Page 30: ...lantate verwenden deren Oberfl che besch digt ist Besch digte Implantate m ssen entsorgt werden Anwender m ssen geeignete pers nliche Schutzausr stung PSA tragen Alle Anwender m ssen qualifizierte Mit...

Page 31: ...sollten abgewischt werden um ein Antrocknen an der Oberfl che zu vermeiden Alle Anwender m ssen qualifizierte Mitarbeiter mit dokumentiertem Nachweis ihrer Schulung und Fachkompetenz sein Die Schulun...

Page 32: ...rs auf schwer zug ngliche Bereiche zu achten Besondere Sorgfalt ist bei kan lierten Instrumenten erforderlich welche mit einer geeigneten Flaschenb rste gereinigt werden m ssen Bei freiliegenden Feder...

Page 33: ...ende Oberfl chen lassen sich mit einer Spritze oder einer Wasserd se besser aussp len 3 Die Instrumente aus der Enzyml sung nehmen und mindestens eine 1 Minute lang mit deionisiertem Wasser absp len 4...

Page 34: ...terility assurance in health care facilities Ausf hrliche Anleitung zur Dampfsterilisation und Sterilit tssicherheit in Gesundheitseinrichtungen zu beachten Anweisungen f r bestimmte Sterilisatoren un...

Page 35: ...sind m glicherweise nicht in allen L ndern zum Gebrauch freigegeben Keine der in diesem Material enthaltenen Aussagen darf als Werbung f r ein Produkt oder f r die Verwendung eines Produktes auf eine...

Page 36: ...PKGI 79 F Effective 03 2019 3 3 3 ASTM F67 ASTM F136 EL 36 EL...

Page 37: ...PKGI 79 F Effective 03 2019 Acumed acumed net EL 37 EL...

Page 38: ...PKGI 79 F Effective 03 2019 MRI MRI1 1 Shellock F G Reference Manual for Magnetic Resonance Safety Implants and Devices 2011 Edition Biomedical Research Publishing Group 2011 Acumed EL 38 EL...

Page 39: ...PKGI 79 F Effective 03 2019 Acumed SUD SUD pH 8 5 1 2 3 4 pH 8 5 EL 39 EL...

Page 40: ...PKGI 79 F Effective 03 2019 1 2 15 3 4 pH 8 5 Acumed 2 2 pH 8 5 5 40 C pH 8 5 2 40 C 40 90 C EL 40 EL...

Page 41: ...PKGI 79 F Effective 03 2019 Acumed pH pH 11 1 2 20 3 3 4 EL 41 EL...

Page 42: ...PKGI 79 F Effective 03 2019 10 5 3 6 7 8 1 2 10 3 1 4 EL 42 EL 1 2 2 20 3 1 4 15 X2 5 2 64 66 C 146 150 F 6 15 7 10 64 66 C 146 150 F 8 7 116 C 240 F...

Page 43: ...PKGI 79 F Effective 03 2019 25 0 kGy Acumed Acumed Acumed Acumed AORN ANSI AAMI ST79 2010 EL 43 EL...

Page 44: ...PKGI 79 F Effective 03 2019 1 1 SAL 2 ANSI AAMI ST79 2010 8 8 1 80 2340 80 2341 3 132o C 270o F 4 30 2 30 EL 44 EL...

Page 45: ...PKGI 79 F Effective 03 2019 EL 45 EL...

Page 46: ...ficiente del hueso y sensibilidad del tejido blando o al material Si se sospecha de sensibilidad al material deber n realizarse pruebas antes del implante Estos dispositivos est n contraindicados en p...

Page 47: ...dos en el procedimiento antes de emplearlo Puede consultar las t cnicas quir rgicas en el sitio web de Acumed acumed net ADVERTENCIAS SOBRE LOS IMPLANTES Para utilizar de forma segura un implante el c...

Page 48: ...antes distintos por motivos de tipo metal rgico mec nico y funcional Las ventajas de la cirug a de implante pueden no satisfacer las expectativas del paciente o verse mermadas con el tiempo lo que har...

Page 49: ...dor enzim tico neutro o detergente neutro con un pH 8 5 1 Prepare una soluci n con agua del grifo tibia y detergente o limpiador Siga las recomendaciones de uso del fabricante del limpiador enzim tico...

Page 50: ...os instrumentos para evitar que queden residuos Con los instrumentos de Acumed no se debe usar aceite mineral ni lubricantes de silicona Para limpiar el instrumental reutilizable se recomiendan limpia...

Page 51: ...a unidad de ultrasonidos con soluci n de limpieza Accione todas las piezas m viles para que el detergente entre en contacto con todas las superficies Realice la limpieza con ultrasonidos del instrumen...

Page 52: ...mpieza del instrumental quir rgico Se puede utilizar el sistema automatizado como un proceso de seguimiento a la limpieza manual Los instrumentos se deben inspeccionar bien antes de su esterilizaci n...

Page 53: ...te las instrucciones por escrito del fabricante de su equipo para conocer las instrucciones del esterilizador y de configuraci n de carga No se recomienda la esterilizaci n por vapor con desplazamient...

Page 54: ...le que los productos no est n aprobados para su uso en todos los pa ses Nada de lo que contenga esta documentaci n deber interpretarse como una promoci n ni solicitud de cualquiera de los productos ni...

Page 55: ...fection active ou latente sepsis masse osseuse de qualit ou de quantit insuffisante et sensibilit des tissus mous ou allergie au mat riau En cas de doute concernant une allergie au mat riau des tests...

Page 56: ...intervention Les techniques op ratoires peuvent tre consult es sur le site Internet d Acumed acumed net AVERTISSEMENTS CONCERNANT LES IMPLANTS pour une utilisation s re et efficace de l implant le ch...

Page 57: ...onctionnelles d conseillent le m lange de composants implantaires de fabricants diff rents Les b n fices de la chirurgie implantaire peuvent ne pas r pondre aux attentes du patient ou diminuer avec le...

Page 58: ...ion base d eau chaude du robinet et de d tergent ou d un agent de nettoyage Suivre les recommandations d utilisation du fabricant de l agent de nettoyage enzymatique ou du d tergent en s attachant tou...

Page 59: ...apercevoir les instruments dans la solution de nettoyage Les agents de nettoyage doivent pouvoir tre facilement limin s des instruments pour viter les d p ts r siduels Ne pas utiliser de lubrifiants b...

Page 60: ...surer que toutes les fentes sont bien nettoy es 3 Retirer les instruments et rincer abondamment l eau courante pendant trois 3 minutes au moins Pr ter une attention toute particuli re aux cannelures e...

Page 61: ...l obtention d un nettoyage et d une d sinfection en profondeur Instructions concernant le nettoyage la d sinfection automatis s Les syst mes de lavage s chage automatis s sont d conseill s en tant que...

Page 62: ...fabricant du mat riel pour obtenir des instructions sp cifiques la configuration de st rilisation et de chargement La st rilisation la vapeur d placement de gravit est d conseill e La st rilisation d...

Page 63: ...tions relatives des produits qui peuvent tre commercialis s ou non dans un pays sp cifique ou tre commercialis s sous diff rentes marques de commerce dans diff rents pays Les produits peuvent tre appr...

Page 64: ...Sistema di placche per caviglia 3 materiale Se si sospetta sensibilit al materiale dell impianto opportuno eseguire dei test prima di procedere con l intervento L uso di questi dispositivi controindi...

Page 65: ...ER L IMPIANTO Per un utilizzo sicuro ed efficace dell impianto il chirurgo deve conoscere lo strumento i metodi di applicazione gli strumenti e la tecnica chirurgica raccomandata per il dispositivo Il...

Page 66: ...ecessitando cos di intervento di revisione per sostituire l impianto o intervenire con procedure alternative Gli interventi chirurgici di revisione nel caso di impianto non sono infrequenti PRECAUZION...

Page 67: ...colare attenzione ai tempi di esposizione alla temperatura alla qualit dell acqua e alla concentrazione 2 Con attenzione lavare a mano l impianto Non usare detergenti abrasivi o spugnette in lana di a...

Page 68: ...esidui Non utilizzare olio minerale o lubrificanti in silicone sugli strumenti Acumed Per la pulizia degli strumenti riutilizzabili si consiglia l uso di agenti enzimatici e detergenti a pH neutro ess...

Page 69: ...asuoni con la soluzione detergente Azionare tutte le parti mobili per permettere al detergente di entrare in contatto con tutte le superfici Sonicare per un minimo di dieci 10 minuti 5 Rimuovere gli s...

Page 70: ...possibile avvalersi di un sistema automatico come metodo addizionale alla pulizia manuale Ispezionare attentamente gli strumenti prima della sterilizzazione per assicurare che siano effettivamente pul...

Page 71: ...e facilities Consultare le istruzioni scritte fornite dal produttore dell apparecchiatura per le istruzioni relative allo specifico sterilizzatore e alla configurazione del carico La sterilizzazione a...

Page 72: ...e nei differenti paesi I prodotti potrebbero non essere tutti approvati per l uso in tutti i paesi Nessun elemento contenuto nei presenti materiali va inteso come intenzione a promuovere o sollecitare...

Page 73: ...estaat dat een pati nt overgevoelig is voor het materiaal moeten voorafgaand aan de implantatie testen worden uitgevoerd Deze implantaten zijn gecontra indiceerd bij pati nten die instructies met betr...

Page 74: ...an relevante publicaties en ervaren collega s te raadplegen aangaande de ingreep De chirurgische technieken kunt u vinden op de website van Acumed acumed net WAARSCHUWINGEN BETREFFENDE IMPLANTATEN Voo...

Page 75: ...alen Door botplaten veelvuldig te buigen kunnen ze verzwakken wat kan leiden tot ontijdig breken van het implantaat en falen van het implantaat Combineren van implantaatcomponenten van verschillende f...

Page 76: ...epassing zijnde richtlijnen en standaarden en het huidig ziekenhuisbeleid te omvatten Handmatige herverwerking Benodigdheden Een zachtharige borstel een neutraal enzymatisch reinigingsmiddel of een ne...

Page 77: ...d om te voorkomen dat die op het oppervlak opdrogen Alle gebruikers dienen bevoegd personeel te zijn met documenten waaruit de juiste opleiding en voldoende bekwaamheid blijkt De opleiding dient ook d...

Page 78: ...aan gecanuleerde instrumenten en reinig met een geschikte flessenborstel Voor blootliggende veren spiralen of flexibele vormdelen giet een ruime hoeveelheid reinigingsmiddel over de tussenruimtes om...

Page 79: ...ng en spoel minimaal n 1 minuut in gede oniseerd water 4 Plaats de instrumenten in een geschikt mandje voor een reinigings desinfectieapparaat en stel het reinigings desinfectieapparaat in op een stan...

Page 80: ...ion and sterility assurance in health care facilities Raadpleeg de schriftelijke instructies van de fabrikant van uw apparatuur voor specifieke instructies voor uw sterilisator en de wijze waarop die...

Page 81: ...stricties Het kan zijn dat de producten niet in alle landen zijn goedgekeurd voor gebruik Niets in dit materiaal dient te worden beschouwd als promotie of een poging tot verkoop van producten of als e...

Page 82: ...lige eller er ute av stand til f lge postoperative pleieinstrukser er kontraindisert for disse Ankelplatesystem 3 enhetene Enhetene er ikke ment skrus fast eller fikseres til bakre elementer pedikler...

Page 83: ...lige bivirkninger av dette implantatet Disse advarslene skal inkludere muligheten for at enheten svikter eller at behandlingen mislykkes som f lge av l s fiksering og eller at enheten l sner slitasje...

Page 84: ...ndig tilheling eller at det brukes overdreven kraft p implantatet under innsetting Implantatmigrasjon og eller at implantatet l sner kan forekomme Metallf lsomhet histologisk allergisk eller negativ r...

Page 85: ...15 minutter 3 Skyll implantatet grundig med deionisert eller destillert vann Bruk deionisert eller destillert vann for siste skylling 4 T rk implantatet med en ren myk lofri klut for unng riper i over...

Page 86: ...i kontakt med alle overflater La ligge i v sken i minst tjue 20 minutter Bruk en myk nylonb rste til skrubbe instrumentene forsiktig inntil all synlig smuss er fjernet V r spesielt oppmerksom p vanske...

Page 87: ...g eventuelle tett motliggende overflater 3 Fjern instrumentene fra enzyml sningen og skyll dem i deionisert vann i minst ett 1 minutt 4 Plasser instrumentene i et egnet vaskeenhet desinfektorkurv og k...

Page 88: ...ization and sterility assurance in health care facilities R df r deg med utstyrsprodusentens skriftlige instrukser for spesifikke instrukser om sterilisator og lastekonfigurasjon Dampsterilisering med...

Page 89: ...ler restriksjoner i ulike land Produktene vil muligens ikke v re godkjent for bruk i alle land Ingenting angitt i denne dokumentasjonen skal oppfattes som en kampanje for eller oppfordring til kj p av...

Page 90: ...m caso de suspeita de sensibilidade ao material devem ser realizados testes antes do implante Os pacientes que n o estejam dispostos ou sejam incapazes de seguir instru es de cuidados p s operat rios...

Page 91: ...antes de utilizar os produtos Poder encontrar as t cnicas cir rgicas no website da Acumed acumed net ADVERT NCIAS RELATIVAS AO IMPLANTE Para uma utiliza o segura e eficaz do implante o cirurgi o deve...

Page 92: ...a utiliza o simult nea de componentes de implante de diferentes fabricantes Os benef cios da cirurgia de implante podem n o cumprir as expectativas do paciente ou podem deteriorar se com o tempo sendo...

Page 93: ...ro com um pH 8 5 1 Prepare uma solu o utilizando gua corrente t pida e detergente ou agente de limpeza Siga as recomenda es de utiliza o do fabricante do detergente ou do agente de limpeza enzim tico...

Page 94: ...ntos para prevenir a acumula o de res duos N o se recomenda a utiliza o de lubrificantes de leo mineral ou de silicone em instrumentos Acumed Recomenda se a utiliza o de agentes de limpeza enzim ticos...

Page 95: ...ultra s nica com solu o de limpeza Desloque todas as pe as m veis para permitir que o detergente entre em contacto com todas as superf cies Coloque os instrumentos no aparelho de ultra sons durante n...

Page 96: ...os como o nico m todo de limpeza para instrumentos cir rgicos Poder ser utilizado um sistema automatizado como processo de seguimento da limpeza manual Os instrumentos devem ser minuciosamente inspecc...

Page 97: ...lity assurance in health care facilities Guia abrangente para a esteriliza o a vapor e garantia de esterilidade em institui es de cuidados de sa de Consulte as instru es do fabricante do equipamento p...

Page 98: ...s indica es ou restri es em diferentes pa ses Os produtos podem n o estar aprovados para utiliza o em todos os pa ses Nada nestes materiais deve ser interpretado como uma promo o ou solicita o de qual...

Page 99: ...ilkan tukilevyj rjestelm 3 herkkyytt materiaalille materiaali tulee testata ennen implantointia Potilaat jotka eiv t halua tai kykene noudattamaan leikkauksenj lkeisi hoito ohjeita ovat my s vasta aih...

Page 100: ...suunniteltu kest m n painon kuorman tai liiallisen aktiviteetin aiheuttamaa rasitusta Laitteen v r sis nvienti implantoinnin aikana saattaa lis t irtoamisen tai siirtymisen mahdollisuutta Potilasta t...

Page 101: ...tit hankautumiselta ja nirhautumiselta sill t m n tyyppinen kuormituksen keskittyminen saattaa ohtaa laitteen toimintah iri n HAITTAVAIKUTUKSET Mahdollisiin haitallisiin vaikutuksiin sis ltyy kipua ep...

Page 102: ...sesti deionisoidulla tai puhdistetulla vedell K yt lopulliseen huuhteluun deioinisoitua tai puhdistettua vett 4 Kuivaa implantti puhtaalla pehme ll ja nukkaamattomalla liinalla jotta pinta ei naarmuun...

Page 103: ...vin t rke ett em ksiset puhdistusaineet neutralisoidaan t ydellisesti ja huuhdotaan pois instrumenteista Kirurgiset instrumentit tulee kuivata perusteellisesti jotta ruostetta ei muodostuisi vaikka ne...

Page 104: ...eiden huuhteluun 6 Tarkasta instrumentit normaalivalossa n kyv n lian varalta 7 Jos instrumenteissa n kyy likaa toista yll kuvatut ultra nik sittely ja huuhteluvaiheet 8 Poista instrumenteista liialli...

Page 105: ...Vaihe Kuvaus 1 Kahden 2 minuutin esipesu kylm ll johtovedell 2 Kahdenkymmenen 20 sekunnin entsyymisuihkutus kuumalla johtovedell 3 Yhden 1 minuutin entsyymiupotus 4 Viidentoista 15 sekunnin huuhtelu...

Page 106: ...Settings ja vaatimuksia jotka l ytyv t oppaasta ANSI AAMI ST79 2010 Komprehensiivinen opas h yrysterilointiin ja SAL arvoihin terveydenhuollon laitoksissa Tutustu laitteen valmistajan kirjallisiin oh...

Page 107: ...k ytt aiheilla tai rajoituksilla eri maissa Tuotteita ei ehk ole hyv ksytty k ytt n kaikissa maissa T m n aineiston sis lt ei tule tulkita mink n tuotteen mainokseksi tai mink n tuotteen k yt n edist...

Page 108: ...a v rdinstruktioner r kontraindicerade f r dessa Plattsystem f r fotled 3 enheter Dessa enheter r inte avsedda f r skruvinf stning eller fixering i de bakre elementen pediklarna i hals br st eller l n...

Page 109: ...kationerna av detta implantat Dessa f rsiktighets tg rder inkluderar risken f r att enheten eller behandlingen misslyckas p grund av l s fixering och eller lossnande p frestning alltf r omfattande akt...

Page 110: ...n intr ffa p grund av alltf r omfattande aktivitet l ngvarig belastning p enheten ofullst ndig l kning eller att stor kraft ut vas p implantatet under inf randet Implantatmigrering och eller lossning...

Page 111: ...atur vattenkvalitet och koncentration 2 Reng r implantaten med ultraljud i minst 15 minuter 3 Sk lj implantatet noga med avjoniserat eller renat vatten Anv nd avjoniserat eller renat vatten f r den si...

Page 112: ...medel och reng ringsmedel med den f rsp dning och vid den temperatur som rekommenderas av tillverkaren F rska l sningar b r beredas n r befintliga l sningar blivit kraftigt kontaminerade 2 Placera in...

Page 113: ...dlig reng ring av instrumenten Anv ndning av en spruta eller vattenstr le f rb ttrar spolningen av sv r tkomliga omr den och eventuella kraftigt avmattade ytor 3 Ta upp instrumenten fr n enzyml sninge...

Page 114: ...sterility assurance in health care facilities Omfattande guide till ngsterilisering och sterilitetss kring p sjukv rdsinstitutioner L s utrustningstillverkarens skriftliga anvisningar f r den specifi...

Page 115: ...sningar i olika l nder Det kan h nda att produkterna inte r godk nda f r anv ndning i alla l nder Ingenting i dessa dokument b r uppfattas som reklam eller intygande om n gon produkt eller anv ndning...

Page 116: ...ulamas ncesinde Ayak Bile i Plaka Sistemi 3 test ger ekle tirilmesi gerekir Postoperatif bak m talimatlar n uygulamak istemeyen veya uygulayamayacak durumda olan hastalar bu cihazlara kontrendikedir B...

Page 117: ...rabilir Hastan n tercihen yaz l olarak bu implant n kullan m hakk ndaki k s tlamalar ve olas advers etkiler konusunda uyar lmas gerekir Bu uyar lar kapsam nda gev ek yap lm sabitleme ve veya gev eme g...

Page 118: ...a r rahats zl k veya anormal hisler ve implant n kendisinden veya cerrahi travmadan kaynaklanan sinir veya yumu ak doku hasar d r A r aktivite cihaza uzun s reli y k uygulanmas iyile me safhas n n ta...

Page 119: ...suyu ve deterjan veya temizleyici kullanarak zeltiyi haz rlay n Do ru ekspoz r s resi s cakl k su kalitesi ve konsantrasyon de erine zellikle nem g stermek i in enzimatik temizleyici veya deterjan re...

Page 120: ...ri kullanmay n Ayr ca pH de eri 11 den y ksek olan zeltilerde anotlama katman z nebilir Manuel Temizlik Dezenfeksiyon Talimatlar 1 Enzimatik maddeleri ve temizlik maddelerini reticinin nerdi i kullan...

Page 121: ...nmesine yard m eder Ula lmas zor b lgelerin ve s k ca kenetlenmi t m y zeylerin daha iyi y kanabilmesi i in bir r nga veya su jetinden faydalan n 3 Aletleri enzim zeltisinden kar n ve en az bir 1 daki...

Page 122: ...ive guide to steam sterilization and sterility assurance in health care facilities i Sa l k bak m tesislerinde sterilizasyon garantisi ve buhar sterilizasyonu i in kapsaml k lavuz uygulay n Belli ster...

Page 123: ...tahliye edilebilir r nler t m lkelerde kullan lmak zere onaylanmayabilir Bu materyallerde bulunan hi bir ey okuyucunun bulundu u lkede ge erli olan yasalar ve d zenlemeler kapsam nda onayl olmayan bel...

Reviews: