Electric Techniques
Surface Potential Detection—Preparation
298
Dimension 3100 Manual
Rev. D
16.5 Surface Potential Detection—Preparation
It is often desirable to apply a voltage to one or more areas of a sample. This may be done in two
ways: by connecting a voltage to the sample mounting chuck, or by making direct contact to the
sample. In both cases, jumper configurations in the back panel of the microscope must be changed
to match the environment desired.
Note:
In addition to any reconfigured jumpers, remember to connect the common or
negative terminal of an external power supply to the Dimension ground or the
AFM chassis.
1. Power down the NanoScope controller and turn off all peripherals. Unplug the NanoScope
power cable from the microscope’s electronic box.
2. Remove the back panel on the microscope’s electronic box.
3. Locate header and jumpers per
on the main electronics backplane.
Note:
Jumper systems without the Extender Electronics Module should appear as
shown in
; whereas jumper systems with the Extender Electronics
option should appear as in
4. Reconfigure jumpers on the backplane header using the appropriate jumper configuration.
5. After correctly configuring the backplane jumpers, replace the cover on the electronics box.
6. Plug the power cable back into the microscope.
7. Turn on the NanoScope controller.
8. Depending upon whether voltage is to be applied to the sample directly or indirectly,
reconfigure jumpers if indicated.
16.5.1 Applying Voltage to the Sample Directly
When voltage is applied directly to the sample, there is no need to reconfigure the jumpers. They
should remain jumpered as shipped from the factory (
), and the sample should be
electrically insulated from the chuck.
Connect the external voltage source directly to the sample by attaching fine gauge wire to
appropriate contacts (e.g., on integrated circuits connect electrical leads directly to pads). For
normal operation, the sample chuck is held at ground. For Application Module-ready systems, the
bias is normally applied to the chuck. Be certain to carefully any electrical connections from the
sample chuck.
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......