Fluid Imaging
Operating Principles
Rev. D
Dimension 3100 Manual
171
10.5.17 Engage
1. Select
Motor
>
Engage
. A pre-engage check begins, followed by Z-stage motor motion.
2. To move to another area of the sample, execute a
Withdraw
command to avoid damaging
the tip and scanner.
3. Select
Stage
>
Focus Surface
and move the stage using the trackball.
Note:
After the tip is engaged, adjust the control panel values to provide the desired
scan parameters.
10.5.18 Adjust Scan Parameters
Once engaged, the scan parameters should be adjusted to obtain the best image. Set the
Slow
scan
axis
to
Disabled
and select
View
>
Scope Mode
. Observe the agreement between the trace and
retrace lines into, and adjust the
Setpoint
,
Gains
and
Scan
rate
to bring the trace and retrace lines
in to coincidence. This procedure will be similar to operation in air, with the following exceptions:
•
Samples are generally softer in fluids. Adjusting the applied forces becomes more
critical.
•
Scan rates tend to be slower in fluid.
Adjusting Setpoint
Contact Mode
1. Adjust the
Setpoint
to as low a value as possible using the cursor keys until the cantilever
pulls off the surface (and the Z-center voltage jumps to -220V).
2. Increase the
Setpoint
slightly until the cantilever begins to touch the surface again and an
image appears. Or, use the
Force
Calibration
command to select the
Setpoint
and estimate
the contact force, as described in
Note:
The cantilever will typically adhere to the sample surface much less in fluid;
therefore, it is often possible to image at much smaller contact forces in liquid
than in air.
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......