Head, Probe, & Sample Preparation
Advanced AFM Operation
Rev. D
Dimension 3100 Manual
117
7.2.16 Engage
•
Select
Motor
>
Engage
, or click the
Engage
icon.
•
Once engaged, refer to the appropriate chapter in this manual for further instructions for
the mode you are using (e.g.
7.2.17 Withdraw
When imaging is complete, withdraw the tip by selecting
Motor
>
Withdraw
, or the click
Withdraw
icon.
This typically raises the tip 1mm above the sample.
Although the 1mm withdraw distance provides enough clearance to loosen the dovetail and remove
the AFM head, if further distance is desired perform one of the following:
•
Select
Stage
>
Focus Surface
to raise the Z-stage with the trackball.
•
Execute
Withdraw
multiple times.
•
Select
Stage
>
Load New Sample
.
Note:
It is impossible to move the Dimension head if the laser sum
signal is too low.
7.3
Advanced AFM Operation
7.3.1 Stage Parameters
The system ships with the following default stage parameters (see
become familiar with using the system, you may want to adjust the stage parameters to speed up the
engage sequence.
1. To access the stage parameters panel, select
SPM Parameters
from the
Stage
pop-down
menu.
2. For initial set-up, use the following values. When done, click
OK
to close the
SPM
Parameters
window.
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......