Facilities Requirements
Acoustic/Vibration Specifications
50
Dimension 3100 Manual
Rev. D
3.5
Acoustic/Vibration Specifications
The following conditions must be met in order to achieve 0.5 angstrom RMS noise specifications:
•
Acoustic: Acoustic noise should not exceed 75dBC (Note “C” weighting).
•
Vibration: Vibration of the SPM mounting surface should not exceed VC-D in any
direction, vertical or horizontal.
Figure 3.5a
Vibration Criteria Plot
Velocity Level in dB re 1 micro-inch/sec
One-Third Octave Band Frequency in Hz
Workshop (ISO)
Residential Day (ISO)
Perception Threshold (ISO)
Operating Theater (ISO)
BBN Criterion A
BBN Criterion B
BBN Criterion C
BBN Criterion D
BBN Criterion E
32,000
16,000
8,000
4,000
2,000
1,000
500
250
125
100
90
80
70
60
50
16
20
25
31.5
40
50
63
80
100
125
160
4
5
6.3
8
12.5
10
Office (ISO)
BBN Criterion A - Probe Test Equipment. 100X Microscopes
BBN Criterion B - 500X Microscopes. Aligners, Steppers to 5
µ
m Geometries
BBN Criterion C - 1000X Microscopes. Aligners, Steppers to 1.5
µ
m Geometries
BBN Criterion D - Steppers, E-Beams to 0.3
µ
m Geometries, CD Inspection Equipment.
Most SEMs to 50,000X
BBN Criterion E - Anticipated Adequate for Future Fabrication and Test Equipment
for Low Submicron Geometries
1959-940-A
40
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......