TappingMode AFM
Troubleshooting
148
Dimension 3100 Manual
Rev. D
9.5
Troubleshooting
9.5.1 Frequency Response Plot
If a peak in the frequency response plot does not appear, perform the following steps:
1. Increase
Sweep width
to the maximum value.
2. Increase the
Drive amplitude
in intervals of 200-300mV until you have reached 2-3V.
3. If the peak still has not appeared, then increase the
Sweep width
by first increasing the
Center Frequency
, then maximizing the
Sweep width
. If there is still no peak on the
response plot, check the laser alignment.
9.5.2 Engaging the Sample
If the engage aborts because the tip is still too far away from the surface, return to
Stage/Focus
Surface
and use the trackball to relocate the surface. After successful re-engaging, a well formed
image appears on the display monitor.
9.5.3 Cantilever Will Not Tune
If the Cantilever Tune fails, check the following:
•
Ensure the laser is properly aligned at the free end of the cantilever (the end furthest
from the substrate).
•
Ensure there is good physical coupling between the probe and the probe holder. Make
sure the groove the probe is held in is free of debris, and that the probe substrate is in
firm contact with the back wall of the groove and is pushed firmly against one side of
the groove.
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......