TappingMode AFM
Basic TappingMode AFM Operation
Rev. D
Dimension 3100 Manual
141
Feedback Controls Panel
•
Set the
Integral gain
to
0.5
and the
Proportional gain
to
0.7
(see
•
Set the
Look Ahead gain
to zero.
Figure 9.2h
Suggested Feedback Controls Settings
Note:
The
Drive frequency
and
Drive amplitude
values were determined during the
Cantilever Tune procedure. It is not necessary to enter a value for the
Amplitude
setpoint
; this will be determined automatically during the engage
process.
9.2.10 Engage
1. Select
Motor
>
Engage
. A pre-engage check begins, followed by Z-stage motor motion.
2. To move to another area of the sample, execute a
Withdraw
command to avoid damaging
the tip and scanner.
3. Move the stage using the trackball to the next area of interest on the sample.
4. Select
Motor
>
Engage
.
Note:
After the tip engages, adjust the control panel values to provide the desired scan
parameters. Refer to
for scan parameter optimization.
5. After engaging, it may be necessary to recheck the Cantilever Tune parameters. Select
View
>
Sweep
>
Cantilever Tune
. In the
Tip Offset
box, enter
200nm
and click
OK
. Adjust the
Drive
frequency
if necessary.
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......