Surface Potential Imaging—Procedure
304
Dimension 3100 Manual
Rev. D
“Potential” signal), it is possible to use the lock-in signal to provide qualitative images that
reflect the sample surface potential. To view the lock-in signal with the reconfigured
Extender box, select the
Data type
=
Phase
.
16.6.1 Troubleshooting the Surface Potential Feedback Loop
The surface potential signal feedback loop can be unstable. This instability can cause the potential
signal to oscillate or become stuck at 10V or -10V. Here are some tips to see if the feedback
loop is working properly with no oscillation:
•
Go into
Scope Mode
and look at the Potential signal. If oscillation noise is evident in
the signal, reduce the FM gains. If oscillations persist even at very low FM gains, try
increasing the
Lift scan height
and/or reducing the
Drive amplitude
until oscillation
stops. If the tip crashes into the surface the Lock-in signal becomes unstable and can
cause the feedback loop to malfunction. Increasing the
Lift height
and reducing the
Drive amplitude
can prevent this problem. Once oscillation stops, the FM gains may be
increased for improved performance.
In
Scope Mode
, if the Potential signal is perfectly flat and shows no noise even with a small Z-
range, the feedback loop is probably stuck at ±10V. (You can verify this by changing the value of
Realtime planefit
to
None
in the
Channel 1
panel.) Reduce the
Scan rate
and watch the display
monitor which indicates the cantilever amplitude. On the topographic trace, the voltage displayed
should be the setpoint selected for the Main scan. On the Potential trace, this voltage drops close to
zero if the cantilever oscillation is being successfully reduced. If the value on the display monitor
instead goes to a large nonzero value, the feedback loop is probably not working properly. In this
case, try reducing the
Drive amplitude
and increasing the
Lift scan height
. It may also be helpful
to momentarily turn the
Interleave mode
to
Disabled
, then back to
Enabled
. Also try reducing any
external voltage that is being applied to the sample to stabilize the feedback loop, then turn the
voltage back up.
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......