Head, Probe, & Sample Preparation
Basic AFM Operation
114
Dimension 3100 Manual
Rev. D
7.2.12 Load the Sample
If it is your first time operating the microscope, we recommend that you first image the calibration
sample (a 10µm-pitch grid of 200nm step height) provided with the instrument.
Small Sample Preparation
Place the calibration sample or other small sample on one of the 1.5cm diameter metal disks used
for sample mounting. The Dimension 3100 SPM ships with several steel sample disks that you may
attach to a magnetic sample holder located in the Dimension Chuck Spare Parts Kit. Also provided
with the instrument are red and white colored double-sided adhesive patches, or “sticky tabs,”
designed to hold the sample chip to the disk.
1. Peel off a sticky tab from the provided sheet, and place it on the steel small sample puck (see
2. Peel off the red-and-white paper to leave a patch of the two-sided adhesive on the steel
sample disk holding the sample chip to the disk.
3. Using tweezers, place the small sample to be imaged firmly on the adhesive (see
).
Note:
Alternatively, you may glue a small sample down to the sample puck using
cyanoacrylate glue (super glue).
4. Place the small sample disk on the magnetic small sample holder.
Note:
The small sample holder provided with the Dimension 3100 is not optimal for
imaging soft magnetic materials due to the magnetic hold-down properties. If
magnetic imaging is intended, use the vacuum chuck.
Note:
For small samples with flat backs or samples mounted on metal discs, use the
vacuum chuck.
Figure 7.2k
Securing Double-sided Tape to the Sample Disk
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......