Fluid Imaging
Operating Principles
Rev. D
Dimension 3100 Manual
167
7. Center the peak on the frequency sweep plot shown on the display monitor using the
Zoom
In
and
Offset
commands after identifying the maximum amplitude peak with the lowest
frequency in the frequency response plot. This peak is usually in the 8-12kHz range.
Note:
The
Offset
command sets the center frequency equal to the cursor position to
shift the plot. The
Zoom In
command decreases the sweep width and shifts the
center frequency value to stretch the plot.
8. Increase the
Drive Amplitude
until the peak appears.
9. Continue to
Zoom In
and center the peak until the peak coincides with the vertical center
line within
10Hz
. The value displayed for center frequency is now used as the resonant
frequency of the cantilever.
Note:
The system works well in TappingMode if the center frequency is at, or below,
the peak in the resonance plot. The center frequency can decrease to the point
where the oscillation amplitude reaches 90 percent of the maximum value.
Operate at a frequency lower than the resonant frequency to avoid shifting the
resonant frequency upon approach of the tip to the surface (see
).
Figure 10.5d
Cantilever Tune Control Panels for Main Controls
024
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......