Fluid Imaging
Operating Principles
168
Dimension 3100 Manual
Rev. D
10. Adjust the
Drive amplitude
until obtaining an
RMS Amplitude
of
0.3
-
0.6V
. This value
has produced good results for protein samples like RNA polymerase and lysozyme.
Note:
In general, larger RMS amplitudes (approximately 2V) work better for taller
samples such as cells.
Figure 10.5e
Cantilever Tune Frequency Sweep
11. Click
OK
. The parameters set in the
Cantilever Tune
control panel appear in the
Realtime
control panel.
12. Click on
CANCEL
to exit the
Cantilever Tune
command and leave the parameters
unchanged.
Cantilever
Response
0.25 V/div
Center Frequency — 18.44 KHz
0.50 KHz/div
Frequency Sweep
Setpoint