Stage System
Stage Menu Commands
Rev. D
Dimension 3100 Manual
83
3. Select the
Run
option. If you are loading and running the program from the beginning, the
screen prompts the operator whether to refocus the screen.
a.
Clicking
Yes
transfers you to the
Focus Surface
panel; you may then make focusing
adjustments to better view the surface.
b. Clicking
No
initiates the programmed move from step #1.
Figure 5.2n
Initial Focus Prompt
Note:
If the program was previously run without finishing (aborted), the screen
requests whether to begin the program sequence at the aborted step.
4. If you wish to run the entire program from its beginning (step #1); click
No
. Otherwise, click
Yes
(see
5. The screen displays the current program step in progress (see
). To initiate the
program sequence from another starting point, use the
Teach Program
panel to remove any
unwanted step(s), and run the program again from the new “step 1” position.
Figure 5.2o
Programmed Move Prompt
5.2.7 Initialize
The
Initialize
option allows the system software to locate the top (positive) limit switches on the
SPM axis and optics axis stages. If the NanoScope computer determines that it is unsure of the
stage position, it will not allow any functions under the
Stage
menu (except
Initialize
) to be
selected.
To initialize the stage, complete the following:
1. Before beginning initialization, verify that the stage is clear of loose items and debris.
2. Select
Stage
>
Initialize
. The dialog box offers two options (see
Stage
Do you want to set the initial focus?
Yes
No
Programmed Move
Program step 5 in progress
Abort
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......