Installation
Connecting the Dimension 3100 System
62
Dimension 3100 Manual
Rev. D
Note:
The computer ships with the network board disabled to avoid error messages
for computers not used on a network.
1. To use the computer on a network, select
My Computer
>
Control
>
Panel
>
System
>
Hardware Profiles
>
Properties
>
Network
and click to uncheck the box. Click
OK
twice
to exit. Plug in your local network cable and restart the computer.
2. Connect extensions to the Computer. See
and
connection information and connection location.
Table 4.4a
Computer Connections
Fig
Ref
Cable
Part
Number
Box
Function
1
AC Power Cable
466-000-004
Power Strip
Computer power
2
Mouse
n/a
Computer
Input from mouse
2
Keyboard
n/a
Computer
Input from keyboard
3
Trackball
860-000-022
Computer
Input from trackball
(COM1)
4
Serial Cable (12')
464-000-010
D3100
Back Panel
Motor control for stage
(COM2)
5
Monitor Cables (2)
n/a
Monitors
Video signals
6
BNC Cable
462-000-002
D3100
Back Panel
Optical image signal from
framegrabber
7
Serial Cable (6')
NanoScope IV
only
n/a
NanoScope
Controller
Fast Scan (COM4)
8
Cable,
RJ45 LAN
n/a
Host
Network connection
9
Cable, 25-pin D
464-000-012
NanoScope
Controller
Computer to controller
electronic interface
10
Serial Cable (6')
464-000-024
Dimension
Controller
Serial communication w/
Dimension controller
(COM3)
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......