Stage System
Stage Menu Commands
84
Dimension 3100 Manual
Rev. D
Figure 5.2p
Stage Initialize/Cancel Prompt
3. To begin initialization, click
Initialize
. As the stage begins a series of motorized movements,
the screen indicates operating status (see
).
Figure 5.2q
SPM Move to Lower Limits Prompt
Note:
The initialization sequence may be aborted at any time by clicking
Pause
(see
Figure 5.2r
Stage Initializing Prompt
4. The microscope Z-axis moves to the top of its travel. When this is achieved, click
Ok
to
continue.
5. The microscope focuses its camera optics to the extent of its travel. When this is achieved,
click
Ok
to continue (see
Figure 5.2s
Optics Move to End of Travel Prompt
6. Next, the camera zoom optics assembly zooms. When the camera zooms in to the limit of its
travel, click
Ok
to continue (see
).
Figure 5.2t
Stage Zoom Prompt