
Magnetic Force Microscopy
Troubleshooting
270
Dimension 3100 Manual
Rev. D
Note:
This step is not necessary for NanoScope IV.
5. Click the
Ok
button when complete.
15.6 Troubleshooting
15.6.1 MFM Image Verification
The procedure described above should produce a good magnetic force gradient image of the
videotape sample. If there is a problem, check that the Interleave Mode is set to Lift, that Interleave
is Enabled and that the
Scan Line
is set to Interleave. Check also that the
Interleave
values of
Drive Amplitude and Drive Frequency are initially set equal to the main
Scan Controls
values.
15.6.2 Saturation in Amplitude Detection
If using amplitude detection, the magnetic force image can saturate (appear completely featureless)
if the
Interleave
Drive Amplitude
is significantly different than the
Drive Amplitude
in the main
scan. Adjust the
Interleave Setpoint
to restore the image.
Note:
The
Interleave Setpoint
has no physical effect in
LiftMode
since there is no
surface feedback during the lift pass.
15.6.3 Optical Interference
When using
Amplitude Detection
, optical interference may sometimes appear in the Lift
(magnetic force gradient) image when imaging highly reflective samples. Optical interference
appears as evenly spaced, sometimes wavy lines with ~1–2µm spacing superimposed on the lift
image. This occurs when ambient laser light (i.e., light passing around or through the cantilever,
then reflecting off the sample) interferes with laser light reflecting from the cantilever. Interference
can be alleviated by moving the beam spot up the cantilever away from the tip; one-third of the
cantilever length works well. The adjustment can be refined by carefully moving the beam spot
laterally a small distance on the cantilever while scanning until interference fringes are minimized.
Be careful not to move the beam off the cantilever or feedback may be lost.
Note:
Optical interference is essentially eliminated by using phase detection or
frequency modulation.
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......