Electric Techniques
Electric Force Microscopy
Rev. D
Dimension 3100 Manual
283
External Voltage Source Applied to the Tip (No Extender Electronics)
In some cases, it may be advantageous to use voltages greater than 12 V, or to use a pulsed power
supply. If an external source of voltage is to be applied to the tip, configure jumpers as shown in
Figure 16.2h
Jumper Configuration for Applying External Voltage to Tip
Place a current-limiting resistor (e.g., 10–100M
Ω
) in series with the external voltage supply as
shown to protect the tip and sample from damage. You may also use current-limited power supplies.
Connect voltage leads to pins on the header using soldered, push-on connectors. Do not solder leads
directly to the header pins, as the heat may cause damage and/or make jumpering the pins difficult.
Sample Chuck Ground/Bias
Analog 2
Gain Select
Auxiliary D (to NanoScope III controller)
Ground
STM Signal (from Dimension head)
Indicates Jumpers
Unused
Ground
Ground
AFM Tip
Sample Chuck
(-)
(+)
External Voltage
Source
>10 MW
Sample
Tip
+
–
V