Stage System
Stage Menu Commands
78
Dimension 3100 Manual
Rev. D
Figure 5.2g
Set Reference Prompt
3. Click
Origin
in the
Set Reference
panel; this is the first of two points to be used for defining
a reference line on the coordinate grid.
4. Quit the
Set Reference
menu, and use the
Focus Surface
option and trackball to either move
the stage right or upward as necessary.
Move Stage Right
a.
Move the stage to the right to a second point on the linear feature. (This second point,
along with the point of origin, defines the X-axis.)
b. Quit the
Focus Surface
option, and return to the
Set Reference
option under the
Stage
pop-down menu.
c.
Click on the
X Axis
option. This establishes the (Y = 0) reference line.
Move Stage Upward
a.
Move the stage upward to a second point on the linear feature. (This second point, along
with the point of origin, defines the Y-axis.)
b. Quit the
Focus Surface
option, and return to the
Set Reference
option under the
Stage
pop-down menu.
c.
Select the
Y Axis
option. This establishes the (X = 0) reference line.
Note:
Accuracy improves if the two points defining a reference line are located at
some distance to each other. For maximum angular accuracy, the two points
should be located at opposite sides of the sample (see
). The left
image depicts two closely located points to define the X-axis. The right image
depicts a more accurate reference line defined by two, widely spaced points.
Figure 5.2h
Defining the X-Axis
Set Reference
X Axis
Y Axis
Origin
Reset
Quit
X
X
XX
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......