Electric Techniques
Electric Force Microscopy
Rev. D
Dimension 3100 Manual
289
External Voltage Source Applied to Sample (With Extender Electronics)
In some cases, it may be advantageous to utilize voltages greater than 12 V, or to utilize a pulsed
power supply. If an external source of voltage is to be applied to the sample, configure jumpers as
shown in
Figure 16.2n
Jumper Configuration for Applying External Voltage to Sample
Place a current-limiting resistor (e.g., 10-100 M
Ω
) in series with the external voltage supply as
shown to protect the tip and sample from damage.You can also use current-limited power supplies.
Connect voltage leads to pins on the header using soldered, push-on connectors. Do not solder leads
directly to the header pins, as the heat may cause damage and make jumpering the pins difficult.
Chuck Ground/Bias
Unused
Auxiliary 2 Output
Ground
STM signal
Indicates jumpers
Unused
Ground
Ground
AFM Tip
Chuck
(+)
(-)
External Voltage
Source
Analog 2 or
Gnd/OSC + DC signal
Sample
+
–
V
Tip
>10 MW