Electric Techniques
Electric Force Microscopy
Rev. D
Dimension 3100 Manual
277
Figure 16.2c
Microscope Backplane
Instructions for Reconfiguring Jumpers
Carefully examine the jumper configuration figures in the following pages and identify which
jumper configuration is correct for your application. If the configuration you choose differs from
the configuration as shipped from the factory, follow the instructions below. Refer to
and
for examples of factory jumper configurations.
1. Turn off the NanoScope controller and unplug the power cable from the microscope’s
electronic box.
2. Remove the back panel on the microscope’s electronic box.
3. Locate header and jumpers per
on the main electronics backplane.
Note:
Jumper systems without the Extender Electronics Module should appear as
shown in
; whereas jumper systems with the Extender Electronics
option should appear as in
.
4. Reconfigure jumpers on the backplane header using the appropriate jumper configuration.
5. After correctly configuring the backplane jumpers, replace the cover on the electronics box.
6. Plug the power cable back into the microscope.
7. Turn on the NanoScope controller.
HEADER &
JUMPERS