Column cover
76
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
Position the metal panels on the pot segment by segment and
install them from inside to outside (4 sets of 2 screws).
3. Slide the column cover up and close it (see chapter 9 on
page 31).
4. Open the column cover at the bottom to establish the power
supply (see chapter 8 on page 30).
5. Establish the internal power supply, close the column cover
and turn on the operating table (see chapter 10 on page 33).
6. Put on the pad.
2.
CAUTION
Risk of material damage
The column keypad broadband cable will be damaged if
the metal panel segments are installed incorrectly. During
installation, make sure that all segments are installed with
the large cutout [5] on the side with the column keypad [6]
(left side from the operating table).
TEST
Perform a function test.