Calibrate lift motor
268
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
24.7
Calibrate lift motor
The motor level position is set using the CAN test center. Follow the
software description from the CAN test center.
1. Establish a communication connection to the CAN test center
(PC).
2. Open the bellows and secure above.
3. Open top column covers and guide downward.
4. Place the gauge #1729967 [1] on the lowest guide
block [2], and slowly move the column downward until the
suspension plate [3] is at the upper end of the gauge. If
needed, raise the column upward a bit so that the gauge can
be inserted. Important: do not travel to the mechanical end
stop.
5. On the "Target value" task card, press the "Set values" button
for "Position offset".
6. Reset (restart) the drive.
7. Check the end positions of the table top.
8. Close the connection to the operating table.
Содержание TruSystem 7000
Страница 1: ...TruSystem 7000 Operating Table Service Manual...
Страница 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Страница 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 305: ......