Cable W107
220
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
Refasten cable W107 from the lift motor up to power supply
socket.
6. Set the cable deflector onto the two screws on the column
base and slide it flush against the power supply unit. Mount the
cable deflector to the power supply unit (2 screws on the top)
and tighten the two screws on the column base.
7. Establish the internal power supply, close the column cover
and turn on the operating table (see chapter 10 on page 33).
8. Put on the pad.
TEST
Perform recheck according to IEC 60601-1; perform function test
and final check.