Cable W175
254
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
8. Remove the subfloor cover [4] from the table base (4 screws).
9. Unplug cable W175 [5] from the control module [7].
10. Unplug the other end of cable W175 [5] from the switch on
the drive unit [6].
11. Note the cable routing and attachment.
Unfasten cable W175 [5] from the switch on the drive unit [6]
up to the control module [7] and remove cable W175.
Assembly
1. Restore the original cable routing.
Route cable W175 in the table base.
2. Plug cable W175 into the socket on the control module (see
circuit diagram on page 295).
3. Plug the other end of cable W175 into the switch (see circuit
diagram on page 295).
4. Refasten cable W175 from the control module up to the
switch.
5. Mount the subfloor cover on the table base (4 screws).
6. Mount the cover above the control module (4 screws).
7. Safety note: The operating table is extremely heavy; take extra
care not to strain your back while righting it. A second person
is needed to right the operating table. Clear the floor in front
Содержание TruSystem 7000
Страница 1: ...TruSystem 7000 Operating Table Service Manual...
Страница 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Страница 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 305: ......