Energy chain
50
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
9. Attach the leads of cable W166/W167 to the plug on the
CAN distributor circuit board (center box) (see circuit diagram
on page 295).
10. Restore the original cable routing. Note the shield support.
Fasten ground wire_24 and cables W166/W167 from the
center box to the end of the energy chain on the inside of the
bar.
11. Attach the cover to the center box (see chapter 12.3 on
page 41).
12. Attach the pad plate to the seat section bars (see chapter 12.2
on page 40).
13. Establish the internal power supply, close the column cover
and turn on the operating table (see chapter 10 on page 33).
14. Put on the pad.
TEST
Perform function test and final check.