Frame (connection ring)
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
79
8. Attach the pad plate to the seat section bars (see chapter 12.2
9. Establish the internal power supply, close the column cover
and turn on the operating table (see chapter 10 on page 33).
10. Put on the pad.
14.4
Frame (connection ring)
Disassembly
1. Prepare the operating table (see chapter 7 on page 29).
2. Switch off the operating table and disconnect the internal
power supply (see chapter 10 on page 33).
3. Grasp the column cover, remove the holding screw from the
Trendelenburg assembly and guide the cover carefully
downward.
4. Open the column cover at the top and slide it down (see
5. Remove the metal panel with the column keypad (see
chapter 14.2 on page 77).
6. Separate the frame pieces [1]/[2] from one another (2
screws [3] from below).
7. Note the original installation position of the frame parts.
Remove the complete frame (connection ring with connection
plate bracket and slider plate) from the drive assembly [4] (2
screws each with washers).
TEST
Perform a function test.
Содержание TruSystem 7000
Страница 1: ...TruSystem 7000 Operating Table Service Manual...
Страница 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Страница 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Страница 305: ......