Option B
140
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
2. Insert the cable adapter plug into the ISM module socket.
3. Secure the cable adapter [5] with a cable tie [6] on the ISM
module. The holder features side notches [4] for the cable tie.
4. Mount the bellows on the frame (see chapter 9 on page 31).
5. Establish the internal power supply, close the column cover
and turn on the operating table (see chapter 10 on page 33).
6. Put on the pad.
7. Replace the sticker with the authorization code on the table
base:
a) Remove old sticker.
b) Remove any adhesive residue from the table base and
degrease.
c) Fix new sticker in the same position.
8. Update the software (see the software description from the
CAN test center).
TEST
Perform recheck according to IEC 60601-1; perform function test
and final check.